Partial contact indentation tonometry for measurement of corneal properties-independent intraocular pressure

Match W L Ko, Leo K K Leung, David C C Lam

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

Inter-individual differences in corneal properties are ignored in existing methods for measuring intraocular pressure IOP, a primary parameter used in screening and monitoring of glaucoma. The differences in the corneal stiffness between individuals can be more than double and this difference would lead to IOP measurement errors up to 10 mmHg. In this study, an instrumented partial-contact indentation measurement and analysis method that can account for inter-individual corneal difference in stiffness is developed. The method was tested on 12 porcine eyes ex vivo and 7 rabbit eyes in vivo, and the results were compared to the controlled IOPs to determine the method's validity. Analyses showed that without corneal stiffness correction, up to 10 mmHg of measurement error was found between the existing approach and the controlled IOP. With the instrumented indentation and analysis method, less than 2 mmHg of differences were founded between the measured IOP and the controlled IOP. These results showed that instrumented partial-contact indentation can effectively account for inter-individual corneal variations in IOP measurement.

Original languageEnglish
Pages (from-to)251-268
Number of pages18
JournalMCB Molecular and Cellular Biomechanics
Volume9
Issue number4
Publication statusPublished - 2012
Externally publishedYes

Fingerprint

Manometry
Intraocular Pressure
Individuality
Glaucoma
Swine
Rabbits

Keywords

  • Cornea
  • Corneal stiffness
  • Glaucoma
  • Indentation
  • Intraocular pressure

ASJC Scopus subject areas

  • Cell Biology
  • Biophysics
  • Molecular Biology
  • Molecular Medicine

Cite this

Partial contact indentation tonometry for measurement of corneal properties-independent intraocular pressure. / Ko, Match W L; Leung, Leo K K; Lam, David C C.

In: MCB Molecular and Cellular Biomechanics, Vol. 9, No. 4, 2012, p. 251-268.

Research output: Contribution to journalArticle

@article{dabd48456f404081a75566ef007e60b5,
title = "Partial contact indentation tonometry for measurement of corneal properties-independent intraocular pressure",
abstract = "Inter-individual differences in corneal properties are ignored in existing methods for measuring intraocular pressure IOP, a primary parameter used in screening and monitoring of glaucoma. The differences in the corneal stiffness between individuals can be more than double and this difference would lead to IOP measurement errors up to 10 mmHg. In this study, an instrumented partial-contact indentation measurement and analysis method that can account for inter-individual corneal difference in stiffness is developed. The method was tested on 12 porcine eyes ex vivo and 7 rabbit eyes in vivo, and the results were compared to the controlled IOPs to determine the method's validity. Analyses showed that without corneal stiffness correction, up to 10 mmHg of measurement error was found between the existing approach and the controlled IOP. With the instrumented indentation and analysis method, less than 2 mmHg of differences were founded between the measured IOP and the controlled IOP. These results showed that instrumented partial-contact indentation can effectively account for inter-individual corneal variations in IOP measurement.",
keywords = "Cornea, Corneal stiffness, Glaucoma, Indentation, Intraocular pressure",
author = "Ko, {Match W L} and Leung, {Leo K K} and Lam, {David C C}",
year = "2012",
language = "English",
volume = "9",
pages = "251--268",
journal = "MCB Molecular and Cellular Biomechanics",
issn = "1556-5297",
publisher = "Tech Science Press",
number = "4",

}

TY - JOUR

T1 - Partial contact indentation tonometry for measurement of corneal properties-independent intraocular pressure

AU - Ko, Match W L

AU - Leung, Leo K K

AU - Lam, David C C

PY - 2012

Y1 - 2012

N2 - Inter-individual differences in corneal properties are ignored in existing methods for measuring intraocular pressure IOP, a primary parameter used in screening and monitoring of glaucoma. The differences in the corneal stiffness between individuals can be more than double and this difference would lead to IOP measurement errors up to 10 mmHg. In this study, an instrumented partial-contact indentation measurement and analysis method that can account for inter-individual corneal difference in stiffness is developed. The method was tested on 12 porcine eyes ex vivo and 7 rabbit eyes in vivo, and the results were compared to the controlled IOPs to determine the method's validity. Analyses showed that without corneal stiffness correction, up to 10 mmHg of measurement error was found between the existing approach and the controlled IOP. With the instrumented indentation and analysis method, less than 2 mmHg of differences were founded between the measured IOP and the controlled IOP. These results showed that instrumented partial-contact indentation can effectively account for inter-individual corneal variations in IOP measurement.

AB - Inter-individual differences in corneal properties are ignored in existing methods for measuring intraocular pressure IOP, a primary parameter used in screening and monitoring of glaucoma. The differences in the corneal stiffness between individuals can be more than double and this difference would lead to IOP measurement errors up to 10 mmHg. In this study, an instrumented partial-contact indentation measurement and analysis method that can account for inter-individual corneal difference in stiffness is developed. The method was tested on 12 porcine eyes ex vivo and 7 rabbit eyes in vivo, and the results were compared to the controlled IOPs to determine the method's validity. Analyses showed that without corneal stiffness correction, up to 10 mmHg of measurement error was found between the existing approach and the controlled IOP. With the instrumented indentation and analysis method, less than 2 mmHg of differences were founded between the measured IOP and the controlled IOP. These results showed that instrumented partial-contact indentation can effectively account for inter-individual corneal variations in IOP measurement.

KW - Cornea

KW - Corneal stiffness

KW - Glaucoma

KW - Indentation

KW - Intraocular pressure

UR - http://www.scopus.com/inward/record.url?scp=84875500266&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84875500266&partnerID=8YFLogxK

M3 - Article

VL - 9

SP - 251

EP - 268

JO - MCB Molecular and Cellular Biomechanics

JF - MCB Molecular and Cellular Biomechanics

SN - 1556-5297

IS - 4

ER -